What are the specifications of mass spectrometers?

 A partial pressure measurement unit is characterized essentially by the following properties (DIN 28 410): 

What is line width resolution?

The line width is a measure of the degree to which differentiation can be made between two adjacent lines of the same height. The resolution is normally indicated. It is defined as: R = M / ΔM and is constant for the quadrupole spectrometer across the entire mass range, slightly greater than 1 or ΔM < 1. 

Often an expression such as “unit resolution with 15% valley” is used. This means that the “bottom of the valley” between two adjacent peaks of identical height comes to 15 % of the height of the peak or, put another way, at 7.5 % of its peak height the line width DM measured across an individual peak equals 1 amu (atomic mass unit); see in this context the schematic drawing in Fig. 4.10. 

gas-analysis-and-mass-spectrometers

Fig 4.10 Line width – 15% valley

What is the mass range of mass spectrometers?

The mass range is characterized by the atomic numbers for the lightest and heaviest ions with a single charge which are detected with the unit. 

What is sensitivity in mass spectrometry?

Sensitivity E is the quotient of the measured ion flow and the associated partial pressure; it is normally specified for argon or nitrogen: 

gas-analysis-and-mass-spectrometers

(4.1)

How to define the smallest detectable partial pressure

The smallest detectable partial pressure is defined as a ratio of noise amplitude to sensitivity: 

gas-analysis-and-mass-spectrometers

Smallest detectable partial pressure ratio (concentration)

The definition is: 
SDPPR = pmin / pΣ (ppm) 
This definition, which is somewhat “clumsy” for practical use, is to be explained using the detection of argon36 in the air as the example: Air contains 0.93 % argon by volume; the relative isotope frequency of Ar40 to Ar36 is 99.6 % to 0.337 %. Thus, the share of Ar36 in the air can be calculated as follows: 

gas-analysis-and-mass-spectrometers
gas-analysis-and-mass-spectrometers

Fig 4.11 Detection of Argon35

Figure 4.11 shows the screen print-out for the measurement. The peak height for Ar36 in the illustration is determined to be 1.5 · 10-13 A and noise amplitude Δ · i+to be 4 · 10-14 A. The minimum detectable concentration is that concentration at which the height of the peak is equal to the noise amplitude. This results in the smallest measurable peak height being 1.5 · 10-13 A/2.4 · 10-14 A = 1.875. The smallest detectable concentration is then derived from this by calculation to arrive at: 

gas-analysis-and-mass-spectrometers

What is the linearity range of mass spectrometers?

The linearity range is that pressure range for the reference gas (N2, Ar) in which sensitivity remains constant within limits which are to be specified (± 10 % for partial pressure measurement devices). 
In the range below 1 · 10-6 mbar the relationship between the ion flow and partial pressure is strictly linear. Between 1 · 10-6 mbar and 1 · 10-4 mbar there are minor deviations from linear characteristics. Above 1 · 10-4 mbar these deviations grow until, ultimately, in a range above 10-2 mbar the ions for the dense gas atmosphere will no longer be able to reach the ion trap. The emergency shutdown for the cathode (at excessive pressure) is almost always set for 5 · 10-4 mbar. Depending on the information required, there will be differing upper limits for use. 

In analytical applications, 1 · 10-6 mbar should not be exceeded if at all possible. The range from 1 · 10-6 mbar to 1 · 10-4 mbar is still suitable for clear depictions of the gas composition and partial pressure regulation (see Fig. 4.12). 

gas-analysis-and-mass-spectrometers

Fig 4.12 Qualitative linearity curve

Information on surfaces and amenability to bake-out

Additional information required to evaluate a sensor includes specifications on the bake-out temperature (during measurement or with the cathode or SEMP switched off), materials used and surface areas of the metal, glass and ceramic components and the material and dimensions for the cathode; data is also needed on the electron impact energy at the ion source (and on whether it is adjustable). These values are critical to uninterrupted operation and to any influence on the gas composition by the sensor itself.

Download Software

Fundamentals of Vacuum Technology 

Download our e-Book "Fundamentals of Vacuum Technology" to discover vacuum pump essentials and processes. 

References

Vacuum symbols

Vacuum symbols

A glossary of symbols commonly used in vacuum technology diagrams as a visual representation of pump types and parts in pumping systems

 

READ MORE

Glossary of units

Glossary of units

An overview of measurement units used in vacuum technology and what the symbols stand for, as well as the modern equivalents of historical units

 

READ MORE

References and sources

References and sources

References, sources and further reading related to the fundamental knowledge of vacuum technology

 

READ MORE

Vacuum symbols

A glossary of symbols commonly used in vacuum technology diagrams as a visual representation of pump types and parts in pumping systems

 

READ MORE

Glossary of units

An overview of measurement units used in vacuum technology and what the symbols stand for, as well as the modern equivalents of historical units

 

READ MORE

References and sources

References, sources and further reading related to the fundamental knowledge of vacuum technology

 

READ MORE